SCIENTIFIC INSTRUMENTATION


SEPARATIONS

 

ExperimentHPLC[]

UNDER DEVELOPMENT

Example applications include: Protein purification, Carbohydrate and oligosaccharides analysis, Polynucleotide separation, trace impurity analysis, Liquid chromatography, Affinity chromatography, Ion-exchange chromatography, Size exclusion chromatography, Reverse phase chromatography, Normal phase chromatography, Chiral chromatography

Analytical HPLC with Photo Diode Array and Fluorescence Detector

Quaternary Pump

 

    • 0.1 to 2.2 mL/min flow rate

    • 7,000 PSI maximum pressure

    • 0.01 inch inner diameter PEEK system tubing

    • Dynamic mixer

    • In-line degasser

    • Titanium pump casing with sapphire pump heads

    • 250 mL to 5 L buffer source bottles/carboys

Detectors

 

    • Absorbance diode array (PDA)

       

        • Deuterium lamp

        • 190 to 500 nm absorbance detection (1.2 nm resolution)

        • Quartz flow cell

           

            • 10 mm pathlength

            • 0.5 µL volume

            • 1,000 PSI maximum pressure

Column Oven

 

    • 20 to 90°C (± 0.1°C) incubation

    • Up to 2 columns connected in series per run

    • 300 mm maximum column length

    • 15 mm maximum column outer diameter

Sample Constraints

 

    • 75 µL sample loop

    • 250 µL autosampler syringe

    • Sample chamber temperature controlled from 5 to 40°C (± 3°C)

    • Up to 48x HPLC/CE vials or 2x SBS style deep well plates per batch

Waters Acuity H Class

Waters Acuity FLR

Analytical HPLC with Evaporative Light Scattering Detector

Quaternary Pump

 

    • 0.1 to 2.2 mL/min flow rate

    • 7,000 PSI maximum pressure

    • 0.01 inch inner diameter PEEK system tubing

    • Dynamic mixer

    • In-line degasser

    • Titanium pump casing with sapphire pump heads

    • 250 mL to 5 L buffer source bottles/carboys

Detectors

 

    • Ultraviolet absorbance

       

        • Deuterium lamp

        • 190 to 700 nm wavelength range

           

            • Fused silica flow cell with 10 mm pathlength

            • 0.5 µL volume

            • 1,000 PSI maximum pressure

        • Up to 80 points/second sampling rate

    • Evaporative Light Scattering (ELS)

       

        • Nitrogen sheath gas

           

            • Flow rate between 20 and 60 PSI

            • Heatable to varying magnitude

        • 2 mL/min maximum flow rate

        • Up to 80 points/second sampling rate

        • Drift tube up to 100°C

        • Tungsten halogen light source

Column Oven

 

    • 20 to 90°C (± 0.1°C) incubation

    • Up to 2 columns connected in series per run

    • 300 mm maximum column length

    • 15 mm maximum column outer diameter

Sample Constraints

 

    • 75 µL sample loop

    • 250 µL autosampler syringe

    • Sample chamber temperature controlled from 5 to 40°C (± 3°C)

    • Up to 48x HPLC/CE vials or 2x SBS style deep well plates per batch

Waters Acuity H Class

Waters Acuity ELS

Semi-Preparative HPLC with UV/Vis Detector

Quaternary Pump

 

    • 0.2 to 8 mL/min flow rate

    • 300 PSI minimum pressure

    • 4,000 PSI maximum pressure

    • 0.01 inch inner diameter PEEK system tubing

    • Static mixer

    • In-line degasser

    • Titanium pump casing with sapphire pump heads

    • 250 mL to 20 L buffer source bottles/carboys

Detectors

 

    • Pump pressure sensor

    • Column temperature sensor

    • Absorbance diode

       

        • Tungsten deuterium lamp

        • 190 to 900 nm monochromator

        • 6 nm bandpass

        • Quartz flow cell

           

            • 10 mm pathlength

            • 11 µL volume

            • 725 PSI maximum pressure

Column Oven

 

    • 5 to 80°C (± 0.1°C) incubation

    • 300 mm maximum column length

    • 25 mm maximum column outer diameter

Fraction Collection

 

    • Absorbance-based fraction collection methods

       

        • Absolute threshold

        • Derivative detection

        • Time based

    • Up to 8x 96-well deep well plate fraction containers per batch

Sample Constraints

 

    • 500 µL sample loop

    • 500 µL autosampler syringe

    • Up to 40x HPLC/CE vials or 2x SBS style deep well plates per batch

Dionex ultimate 3000 with variable wavelength detector

Semi-Preparative HPLC with pH/Conductivity, UV/Vis, and Fluorescence Detectors

 

    • 0.2 to 8 mL/min flow rate

    • 300 PSI minimum pressure

    • 4,000 PSI maximum pressure

    • 0.01 inch inner diameter PEEK system tubing

    • Static mixer

    • In-line degasser

    • Titanium pump casing with sapphire pump heads

    • 250 mL to 20 L buffer source bottles/carboys

Quaternary Pump

Detectors

 

    • Pump pressure sensor

    • Column temperature sensor

    • Absorbance diode

       

        • Tungsten deuterium lamp

        • 190 to 900 nm monochromator

        • 6 nm bandpass

        • Quartz flow cell

           

            • 10 mm pathlength

            • 11 µL volume

            • 725 PSI maximum pressure

    • Fluorescence

       

        • 8 µL flow cell volume

        • 290 PSI maximum pressure

        • 200 to 880 nm for excitation light (xenon flash lamp source)

        • 220 to 900 nm for emission light

Column Oven

 

    • 5 to 80°C (± 0.1°C) incubation

    • 300 mm maximum column length

    • 25 mm maximum column outer diameter

Fraction Collection

 

    • Absorbance-based fraction collection methods

       

        • Absolute threshold

        • Derivative detection

        • Time based

    • Up to 8x 96-well deep well plate fraction containers per batch

Sample Constraints

 

    • 500 µL sample loop

    • 500 µL autosampler syringe

    • Up to 40x HPLC/CE vials or 2x SBS style deep well plates per batch

Dionex Fluorescence detector

Dionex ultimate 3000

Dionex PH/Conductivity Detector

Semi-Preparative HPLC with MALS, DLS, and RI Detector

Quaternary Pump

 

    • 0.2 to 8 mL/min flow rate

    • 300 PSI minimum pressure

    • 4,000 PSI maximum pressure

    • 0.01 inch inner diameter PEEK system tubing

    • Static mixer

    • In-line degasser

    • Titanium pump casing with sapphire pump heads

    • 250 mL to 20 L buffer source bottles/carboys

Detectors

 

    • Pump pressure sensor

    • Column temperature sensor

    • Absorbance diode

       

        • Tungsten deuterium lamp

        • 190 to 900 nm monochromator

        • 6 nm bandpass

        • Quartz flow cell

           

            • 10 mm pathlength

            • 11 µL volume

            • 725 PSI maximum pressure

    • Multi-angle light scattering

       

        • Up to 36.6 points/sec sampling rate

        • 200 Da to 1,000 MDa molar mass range

        • 0.5 to 500 nm radius of gyration measurement range

        • Light scattering detectors at 18 angles

        • 658 nm laser wavelength

        • Operating temperature between -15 and 150°C

    • Dynamic light scattering

       

        • Up to 36.6 points/sec sampling rate

        • 0.5 to 30 nm hydrodynamic radius measurement range

    • Refractive Index detector

       

        • 7.4 µL flow cell volume

        • Absolute Refractive Index (RI) measurement mode and Differential Refractive Index (DRI) measurement mode available

        • 1.2 to 1.8 RIU Absolute Refractive Index range

        • 0.002 RIU Absolute Refractive Index sensitivity

        • -0.0047 to +0.0047 RIU Differential Refractive Index range

        • ± 7.5 x 10e-10 RIU Differential Refractive Index sensitivity

        • Temperature range 4 and 65°C

        • 660 nm light source

        • Dynamic Range > 16,000,000

Column Oven

 

    • 5 to 80°C (± 0.1°C) incubation

    • 300 mm maximum column length

    • 25 mm maximum column outer diameter

Fraction Collection

 

    • Absorbance-based fraction collection methods

       

        • Absolute threshold

        • Derivative detection

        • Time based

    • Up to 8x 96-well deep well plate fraction containers per batch

Sample Constraints

 

    • 500 µL sample loop

    • 500 µL autosampler syringe

    • Up to 40x HPLC/CE vials or 2x SBS style deep well plates per batch

Wyatt Dawn w/WYATTQELS Module

Dionex Ultimate 3000

Wyatt Optilab

Semi-Preparative HPLC with MALS, DLS, and RI Detector

Quaternary Pump

 

    • 0.2 to 8 mL/min flow rate

    • 300 PSI minimum pressure

    • 4,000 PSI maximum pressure

    • 0.01 inch inner diameter PEEK system tubing

    • Static mixer

    • In-line degasser

    • Titanium pump casing with sapphire pump heads

    • 250 mL to 20 L buffer source bottles/carboys

Column Oven

 

    • 5 to 80°C (± 0.1°C) incubation

    • 300 mm maximum column length

    • 25 mm maximum column outer diameter

Detectors

 

    • Absorbance diode

       

        • Tungsten deuterium lamp

        • 190 to 900 nm monochromator

        • 6 nm bandpass

        • Quartz flow cell

           

            • 10 mm pathlength

            • 11 µL volume

            • 725 PSI maximum pressure

    • Multi-angle light scattering

       

        • Up to 36.6 points/sec sampling rate

        • 200 Da to 1,000 MDa molar mass range

        • 0.5 to 500 nm radius of gyration measurement range

        • Light scattering detectors at 18 angles

        • 658 nm laser wavelength

        • Operating temperature between -15 and 150°C

    • Refractive Index detector

       

        • 7.4 µL flow cell volume

        • Absolute Refractive Index (RI) measurement mode and Differential Refractive Index (DRI) measurement mode available

        • 1.2 to 1.8 RIU Absolute Refractive Index range

        • 0.002 RIU Absolute Refractive Index sensitivity

        • -0.0047 to +0.0047 RIU Differential Refractive Index range

        • ± 7.5 x 10e-10 RIU Differential Refractive Index sensitivity

        • Temperature range 4 and 65°C

        • 660 nm light source

        • Dynamic Range > 16,000,000

    • Different viscometer detector

       

        • Minimum detectable sample at 0.1 µg of 100 kDa polystyrene in THF at 1 mL/min

        • 6.9 kPa differential pressure range

        • 13,500:1 dynamic range

        • Up to 1.5 mL/min for aqueous sample and 3 mL/min for THF

        • 4 ºC to 70 ºC temperature control

Fraction Collection

 

    • Absorbance-based fraction collection methods

       

        • Absolute threshold

        • Derivative detection

        • Time based

    • Up to 8x 96-well deep well plate fraction containers per batch

Sample Constraints

 

    • 500 µL sample loop

    • 500 µL autosampler syringe

    • Up to 40x HPLC/CE vials or 2x SBS style deep well plates per batch

Wyatt Vicostar

Large Scale Preparative HPLC with UV/Vis Detector

Binary Pump

 

    • Up to 50 mL/min flow rate

    • 6,000 PSI maximum pressure

    • In-line degasser

    • 2 to 20 L buffer source bottles/carboys

Column Selector

 

    • Ambient temperature

    • Switchable for up to 6 columns

Detectors

 

    • Pump pressure sensor

    • Column temperature sensor

    • Ultraviolet absorbance detector

    • Deuterium lamp

    • 190 to 600 nm wavelength range (up to two channels)

    • 1 nm wavelength bandpass

    • Up to 120 samples/second measurement (25 samples/second for dual channel) – Quartz flow cell 3 mm pathlength – 4 µL volume – 1,760 PSI maximum pressure

Fraction Collection

 

    • Absorbance-based fraction collection methods

       

        • Absolute threshold

        • Derivative detection

        • Time based

    • Up to 8x 96-well deep well plate fraction containers per batch

Sample Constraints

 

    • 500 µL sample loop

    • 500 µL autosampler syringe

    • Up to 40x HPLC/CE vials or 2x SBS style deep well plates per batch

1290 Infinity 2 Autoscale Preperative LC System

Analytical HPLC with Refractive Index Detector

Semi-Preparative HPLC with UV/Vis Diode Array + Access for 3rd party detectors

Semi-Preparative HPLC with UV/Vis Diode Array + Fluorescence Detectors

Semi-Preparative HPLC with UV/Vis Diode Array Detector

Binary Pump

 

    • Up to 50 mL/min flow rate

    • 6,000 PSI maximum pressure

    • In-line degasser

    • 2 to 20 L buffer source bottles/carboys

Column Selector

 

    • Ambient temperature

    • Switchable for up to 6 columns

Detectors

 

    • Pump pressure sensor

    • Column temperature sensor

    • Ultraviolet absorbance detector

    • Deuterium lamp

    • 190 to 600 nm wavelength range (up to two channels)

    • 1 nm wavelength bandpass

    • Up to 120 samples/second measurement (25 samples/second for dual channel) – Quartz flow cell 3 mm pathlength – 4 µL volume – 1,760 PSI maximum pressure

Fraction Collection

 

    • Absorbance-based fraction collection methods

       

        • Absolute threshold

        • Derivative detection

        • Time based

    • Up to 8x 96-well deep well plate fraction containers per batch

Sample Constraints

 

    • 500 µL sample loop

    • 500 µL autosampler syringe

    • Up to 40x HPLC/CE vials or 2x SBS style deep well plates per batch

 

 

 

 

 

 

 

 

 

 

Agilent – 1260 Infinity II

Analytical HPLC with Refractive Index Detector

Binary Pump

 

    • Up to 50 mL/min flow rate

    • 6,000 PSI maximum pressure

    • In-line degasser

    • 2 to 20 L buffer source bottles/carboys

Column Selector

 

    • Ambient temperature

    • Switchable for up to 6 columns

Detectors

 

    • Pump pressure sensor

    • Column temperature sensor

    • Ultraviolet absorbance detector

    • Deuterium lamp

    • 190 to 600 nm wavelength range (up to two channels)

    • 1 nm wavelength bandpass

    • Up to 120 samples/second measurement (25 samples/second for dual channel) – Quartz flow cell 3 mm pathlength – 4 µL volume – 1,760 PSI maximum pressure

Fraction Collection

 

    • Absorbance-based fraction collection methods

       

        • Absolute threshold

        • Derivative detection

        • Time based

    • Up to 8x 96-well deep well plate fraction containers per batch

Sample Constraints

 

    • 500 µL sample loop

    • 500 µL autosampler syringe

    • Up to 40x HPLC/CE vials or 2x SBS style deep well plates per batch

 

 

 

 

 

 

 

 

 

 

Agilent – 1290 II

Semi-Preparative HPLC Viscometer

Differential Viscosity  

 

    • Sensitivity (minimum detectable sample):  0.1 μg of 100 kDa polystyrene in THF at 1 mL/min, under typical GPC conditions

    • Differential pressure noise: < 0.05 Pa

    • Differential pressure range: 6.9 kPa, protected against overpressure

    • Dynamic range: 135,000:1

    • Linearity:  0.5% of full scale

    • Differential pressure drift: < 2.5 Pa/hr

    • Capillary bridge autobalance:  Thermal, leak-free mechanism

    • Sample shear rate: 5000 Hz with water at 1 mL/min

Fluidics           

 

    • Max. flow rate: Aqueous: 1.5 mL/min; THF: 3 mL/min

    • Delay volume options: 16.2, 10.8 or 5.4 mL, with additional factory-installed option for 32.4 mL 1

    • Max. permissible pressure inlet to outlet: 690 kPa

    • Wetted Materials:  316 stainless steel, PTFE, zirconium silicate, Kalrez ® , Inconel, 26-1 Stainless Steel

Temperature Control 

 

    • Range:  4 ºC to 70 ºC

    • Stability:  <0.005°C

    • Ambient Range:  0 ºC to 40 ºC, 20% to 70% RH

Electronics     

 

    • Digital Output 20 bits @ 10 Hz

    • Auxiliary I/O   

       

        • Inputs:  Autoinject contact closure, Alarm In, Analog In: 2 differential inputs, 18 bits resolution, range -10 V to +10 V

        • Outputs: Autoinject Retransmit, Alarm Out, Recycle Out

    • On-board Computer  

       

        • Communications         Ethernet

    • Front Panel Display: 7″ (178 mm), high-resolution, multi-touch display for viewing/setting signals, instrument settings and diagnostics.

    • Dimensions:  58 cm (L) x 36 cm (W) x 18 cm (H)

 

 

 

 

 

 

 

 

 

 

Wyatt – Vicostar


ExperimentGasChromatography[]

Gas Chromatography System

Flow System

 

    • He carrier gas available

    • Inlets available:

       

        • Split/Splitless capillary column inlet

        • Multimode temperature-programmable capillary column inlet

        • Column head pressure up to 100 PSIG

Column Oven

 

    • Temperature range: 30 to 450ºC

    • Up to 120ºC/min temperature ramp

    • Up to 2 capillary columns

    • 0.53 mm maximum column inner diameter

Split/Splitless Capillary Inlet

 

    • Suitable for capillary columns up to 0.53 mm inner diameter

    • Split ratio up to 7,500:1 to prevent column overload

    • Up to 1,250 mL/min total flow of He

    • Inlet temperature 30 to 400ºC

Multimode Capillary Inlet

 

    • Suitable for capillary columns up to 0.53 mm inner diameter

    • Split ratio up to 7,500:1 to prevent column overload

    • Up to 1,250 mL/min total flow of He

    • Inlet temperature 30 to 400ºC

    • Temperature programmable with ramp rates up to 900ºC/min

Mass Spectrometry

 

    • Single quadrupole mass spectrometer

    • 1.6 to 1,050 m/z mass range

    • 100 to 200ºC quadrupole temperature

    • 100 to 300ºC ion source temperature

    • 100 to 350ºC transfer line temperature

    • Scan speeds up to 12,500 m/z per second

    • Electron ionization:

       

        • Dual filaments

    • Chemical ionization:

       

        • Single filament

        • Methane reagent gas

        • Positive and negative mode

Flame Ionization Detector

 

    • Responds to most organic compounds

    • Minimum detectable level (for tridecane): < 1.2 pgC/s

    • Linear dynamic range: > 10⁷ (± 10%)

    • Up to 1,000 points/second data measurement

Autosampler Rail System

 

    • Vortex mixing up to 2,000 RPM

    • Sample incubation 30 to 200ºC

    • Liquid, headspace, and solid-phase microextraction sample preparation available

    • 0.01 to 100 µL injection volumes for liquid injection

    • 25 to 2,500 µL injection volumes for headspace injection

    • Headspace syringe heating 40 to 150ºC

Agilent 8890 CS System


ExperimentIonChromatography[]

Ion Chromatography System

Pumps

 

    • Dual channel system

       

        • Anion Channel: Isocratic pump with pure water being primary solvent

        • Cation Channel: Quaternary pump

    • 0.1 to 10 mL/min flow rate

    • 5,000 PSI maximum pressure

    • Static mixer

    • In-line degasser

Eluent Generator

 

    • Electrolytic eluent generation for anionic applications

    • 0.1 to 100 mM concentration

Dynamically Regenerated Suppressors

 

    • Constant voltage and constant current mode

    • Maximum suppressor capabilities:

       

        • Anion channel: 50 μeq/min

        • Cation channel: 35 μeq/min

    • 0.25 to 0.75 mL/min flow rate

Conductance Detectors

 

    • Passivated 316 stainless steel electrode

    • Maximum range up to 15,000 μS conductance measurement

    • 0.7 µL flow cell volume

    • One detector for each channel

    • Shared detection oven with 15 to 60ºC (± 0.5ºC) incubation

Column Oven

 

    • 10 to 70°C (± 0.1°C) incubation

    • One analytical column for each channel

Sample Constraints

 

    • 10 µL sample loops

    • Up to 40x HPLC/CE vials or 2x SBS style deep well plates per batch

    • Sample chamber temperature controlled from 4 to 60°C

Dionex ICS-6000 HPIC System

Ion Chromatography System with Electrochemical Detector

Quaternary Pump

 

    • Single channel system

    • 0.1 to 10 mL/min flow rate

    • 5,000 PSI maximum pressure

    • Static mixer

    • In-line degasser

Ultraviolet absorbance detector

 

    • Tungsten and Deuterium lamp

    • 190 to 900 nm wavelength range (up to four channels)

    • 6 nm wavelength bandpass

    • Up to 100 samples/second measurement

    • PEEK flow cell

       

        • 10 mm path length

        • 11 μL flow cell volume

Electrochemical detector

 

    • < 0.2 μL internal volume

    • 100 PSI maximum pressure

    • Gold, platinum, carbon and silver working electrodes

    • pH-Ag/AgCl reference electrode

Column Oven

 

    • 10 to 70°C (± 0.1°C) incubation

Sample Constraints

 

    • 25 μL sample loop

    • Up to 40x HPLC/CE vials or 2x SBS style deep well plates per batch

    • Sample chamber temperature controlled from 4 to 60°C

Dionex ICS-6000 HPIC System

Dionex eletrochemical detector


ExperimentFPLC[]

Example applications include: Aqueous mobile phase, Protein purification, Biomolecule purification, Fast protein liquid chromatography, Medium-pressure chromatography, Liquid chromatography, Affinity chromatography

Fast Protein Liquid Chromatography System Medium and High Flow

System Pump & Mixer

 

    • 0.001 to 25 mL/min flow rate (50 mL/min for column packing) for Avant 25

    • 0.001 to 150 mL/min flow rate (300 mL/min for column packing) for Avant 150

    • 20 MPa maximum pressure for Avant 25

    • 5 MPa maximum pressure for Avant 150

    • 1.4 L mixer volume

    • ± 0.6% gradient composition accuracy for Avant 25

    • ± 0.8% gradient composition accuracy for Avant 150

Detectors

 

    • Pump pressure sensor

       

        • 0 to 20 MPa range

        • ± 0.02 MPa accuracy

    • Temperature monitor

       

        • ± 1.5°C temperature accuracy

    • Conductance sensor

       

        • 0.01 mS/cm to 1 S/cm conductance range

        • 22 µL flow cell volume

    • Absorbance monitor

       

        • 190 to 700 nm wavelength range in increments of 1 nm wavelength

        • Up to 3 simultaneous wavelengths

        • UV flow cell

           

            • 0.5 mm, 2 mm, or 10 mm pathlengths

            • 1 μL, 2 μL, or 8 μL volumes

            • Noise < 0.08 mAU

    • pH monitor

       

        • 0 to 14 range

        • ± 0.1 pH unit (within pH 2 to 12)

Fraction Collection

 

    • Absorbance-based or conductivity-based fraction collection methods

       

        • Absolute threshold

        • Derivative detection

        • Time based

        • Start collection with sample application or within the gradient

    • 50 mL container, 15 mL container, or 96-well deep well plate fraction collection

Sample Constraints

 

    • Autosampler mode

       

        • 50 to 10,000 µL sample loop volume

        • 1,000 to 10,000 µL autosampler syringe

        • Up to 168x HPLC/CE vials or 2x SBS style deep well plate per batch

        • Up to 12 large sample vials (10 mL) for preparative system

    • Flow injection mode

       

        • Up to 50 mL to 4 L sample container volumes

        • Up to 5 containers

        • Controllable sample introduction rate (up to 150 mL/min)

    • Superloop mode

       

        • 10 mL loop volume

        • Up to 50 mL to 4 L sample container volumes

        • Up to 5 containers

AKTA AVANT 25 and 150


ExperimentFlashChromatography[]

Flash Chromatography System

Binary Pump

 

    • 500 to 200 mL/min flow rate

    • 200 PSI maximum pressure

    • Up to 20 L drum solvent sources

    • Programmable gradient (± 1% accuracy)

Absorbance Detector

 

    • Deuterium lamp

    • 200 nm and 780 nm UV detection (5 nm bandpass)

    • 0.6 µl flow cell (0.1 mm pathlength)

Fraction Collection

 

    • 30x 50 mL vials per batch

    • 70x 25 mL vials per batch

Column Constraints

 

    • Luer lock connector pre-packed columns

    • Up to 70 mm outer diamter flash columns

    • Up to 16.5″ length flash columns

    • ~5 to ~660 g of resin per column

Sample Constraints

 

    • Up to 33 g of solid material for dry loading

    • Up to 50 mL of liquid material for wet loading

Teledyne combiflash RF 200


ExperimentPAGE[]

ExperimentAgaroseGelElectrophoresis[]

Robotic Gel Electrophoresis System

Polyacrylamide Gel Electrophoresis

 

    • Slab Gels

       

        • 4.5% and 10% TBE-Urea gels

        • 10% TBE gels

        • 7% and 12% Tris-Glycine gels

        • 5 mm width lanes

        • 3 mm depth gels

    • Electrophoresis

       

        • 0 to 125 V

        • 1 to 100% duty cycle options

    • Fluorescence Imaging

       

        • Post stained gels (stained with any custom solution)

        • Excitation filters

           

            • 470 nm (70 nm bandpass)

            • 625 nm (70 nm bandpass)

        • Emission filters

           

            • 540 nm (40 nm bandpass)

            • 678 nm (44 nm bandpass)

    • Sample Constraints

       

        • Up to 20 µL of sample and loading buffer per lane

        • Up to 10 lanes per gel

        • Up to 4 gels per batch

Agarose Gel Electrophoresis

 

    • Gels

       

        • Analytics (no sample recovery) or size selection (targeted recovery of analytes of particular sizes)

        • 0.5, 1.0, 1.5, 2.0, and 3.0% agarose

        • Independent lanes

    • Electrophoresis

       

        • 100 V

        • 1 to 100% duty cycle options

    • Fluorescence Imaging

       

        • Pre-stained markers ranging from 25 to 20,000 base pairs

        • Multiple in-run images

        • Excitation filters

           

            • 470 nm (70 nm bandpass)

            • 625 nm (70 nm bandpass)

        • Emission filters

           

            • 540 nm (40 nm bandpass)

            • 678 nm (44 nm bandpass)

    • Sample Constraints (Size Selection)

       

        • Up to 50 µL of sample and loading buffer per lane

        • Up to 12 lanes per gel

        • Up to 4 gels per batch

    • Sample Constraints (Analytical)

       

        • Up to 8 µL of sample and loading buffer per lane

        • Up to 23 lanes per gel

        • Up to 4 gels per batch


ExperimentWestern[]

ExperimentTotalProteinDetection[]

Example applications include: Capillary-based Analytical Assay, Gel Electrophoresis, Capillary Gel Electrophoresis, Total Protein Separation, Total Protein Visualization, Protein Size Characterization, Protein Abundance Comparison

Capillary Western Blot System

Capillary Gel Electrophoresis

 

    • 2 to 440 kDa molecular weight range

    • 0 to 500 V separation voltage

    • UV capillary crosslinking for protein immobilization

Secondary Antibodies Available

 

    • Secondary Antibodies Available

       

        • Goat Anti-Rabbit HRP

        • Goat Anti-Mouse HRP

        • Goat Anti-Human IgG HRP

        • Donkey Anti-Goat HRP

CCD Camera-Based Capillary Imaging

 

    • 100 ms to 512 sec exposure times

    • Up to picogram sensitivity (depending on the primary antibody sensitivity)

Capacity

 

    • 13 or 25 capillaries per batch

    • 5 to 8 µL of sample required per capillary

ProteinSimple Jess


ExperimentCrossFlowFiltration[]

Example applications include: Ultrafiltration, Microfiltration, Diafiltration, Protein Purification, Tangential Flow Filtration, Buffer Exchange, Desalting

Tangential Flow Filtration System

 

    • Volume range of 10 mL to 2 L

    • Filtration area of 13 to 3,700 cm²

    • Digitally controlled peristaltic pump

    • Digital pressure monitor

    • Automatic backpressure valve

    • Digital conductivity monitors

    • Absorbance monitor of 200 to 800 nm

    • UV-Vis spectrometer

       

        • 200 to 1,100 nm wavelength range

        • CCD detector allows the entire wavelength

        • Sub-nanometer resolution

        • Up to 200 measurements/sec

 

 

 

 

 

 

 

 

 

 

µPULSE TFF system

Ocean Insight Flame UV-VIS Spectrometer


ExperimentDialysis[]

Example applications include: Fragile Proteins, Viscous Fluids, Polymer Gels, Equilibrium Dialysis, Small Molecule Binding to Proteins, Purification, Buffer Exchange, Desalting, Protein Purification

Static Dialysis

Beakers Available

 

    • 250 mL, 600 mL, 1 L, and 2 L

    • Incubation range 5 to 45°C (below ambient temperatures available for 250 mL beakers only)

    • Buffer mixing available

Tube-Based Dialysis Kits

 

    • 3.5k, 6 to 8k and 12 to 14k molecular weight cutoffs (MWCOs) available

    • 10 µL to 20 mL sample volume range

    • 4 to 60°C incubation range

    • Up to 9 samples per batch

Dialysis Tubing

 

    • 3.5k, 6 to 8k, and 12 to 14k MWCOs available

    • 30 to 130 mL sample volume range

    • Continuous buffer flow to increase separation efficiency

    • 1 sample per batch

Equilibrium Dialysis

 

    • 8 MWCO available

    • 50 to 500 µL sample volume range

    • 4 to 60°C incubation range

    • Up to 48 samples per batch

Tube based dialysis kit

Dialysis Tubing

Equilibrium Dialysis

Dynamic Dialysis

 

    • 3.5k, 6k, and 12k MWCOs available

    • 30 to 130 mL sample volume range

    • Continuous buffer flow to increase separation efficiency

    • > 90% protein recovery after dialysis

    • 1 sample per batch

Spectraflo Lab System


ExperimentThinLayerChromatography[]

High Performance TLC System (HPTLC)

Derivatization

 

    • Automated micro-droplet spraying of chemical stains

    • Application of aqueous or organic solutions

    • Spot-to-spot relative standard deviation <5%

Application

 

    • Up to 15 samples per HPTLC plate (20 x 10 cm)

    • 5 to 6µm silica particle size

    • Samples applied to plates as 8 mm bands for HPTLC

    • Samples applied as spots for reaction monitoring

Development

 

    • Gas phase can be actively controlled to ensure saturation of elution chamber

    • Solvent front monitoring and automatic plate removal

    • Development of TLC and HPTLC plate up to 20×20 cm

Visualization

 

    • UV (254 and 366 nm) and white light illumination

    • Visualize TLC and HPTLC plates up to 20×20 cm

    • Rf evaluation and comparison within or between plates

    • Semi-quantitative peak integration

Camag Derivatizer

Cama Automatic TLC Sampler 4 (ATS 4)

Camag Automatec Developing Chamber 2

Camag TLC Visualizer 2


ExperimentLiquidExtraction[]

Phase Separator

 

    • Rapidly separates immiscible mixtures of organic and aqueous solutions via filtration

    • Separation does not depend on whether the aqueous layer is on the top or bottom

    • Up to 70 ML capacity per filtration

    • Organic solvent flow rate up to 15 mL/min

Separatory Funnel

 

    • Collects aqueous and organic layers of an immiscible mixture at a volume range of 10 mL to 1 L

Robotic Extraction

 

    • Separate top layer from bottom layer given exact height of the top and bottom layer

    • Control aspiration flow rate

Phase Separator

Separatory Funnel

Gilson GX-271

Hamilton Star


ExperimentMagneticBeadSeparation[]

Example applications include: Magnetic Cell Separation, Protein Purification, Immunoprecipitation, Nucleic Acid Isolation

Magnetic Racks

Robotic Separation

 

    • Performed by a Hamilton liquid handler with the Alpaqua Magnum FLX magnet

    • Sample volume constraints

       

        • 96-well plate magnet: up to 2 mL

Manual Separation

 

    • Performed by an operator with the DynaMag magnets

    • Sample volume constraints

       

        • 96-well plate magnet: up to 0.2 mL

        • 2 mL tube magnet: up to 2 mL

        • 15 mL tube magnet: up to 15 mL

        • 50 mL tube magnet: up to 50 mL