SCIENTIFIC INSTRUMENTATION


SPECTROSCOPY

ExperimentNMR[]

Example applications include: 1D NMR spectra for various nuclei (1H, 13C, 19F, 31P) and Water suppression techniques (WATERGATE, Excitation sculpting, Presaturation)

ExperimentNMR2D[]

Example applications include: Correlation Spectroscopy (COSY), Double Quantum Filtered COSY (DQF-COSY), COSY-β, Total Correlation Spectroscopy (TOCSY), Heteronuclear Multiple Bond Correlation (HMBC), Heteronuclear Single Quantum Coherence (HSQC), HSQC-TOCSY, Heteronuclear Multiple Quantum Coherence (HMQC), HMQC-TOCSY, Nuclear Overhauser Effect Spectroscopy (NOESY), Rotational Frame Nuclear Overhauser Effect Spectroscopy (ROESY)

500 MHz NMR

11.74 T Magnet

  • 500 MHz for ¹H
  • 126 MHz for ¹³C
  • 202 MHz for ³¹P
  • 470 MHz for ¹⁹F

Probe

  • SmartProbe ¹H/¹⁹F/¹³C/³¹P 5 mm probe (-150°C to 150°C)
  • Inverse Triple Resonance (TXI) ¹H/¹³C/¹⁹N 5 mm probe (-150°C to 150°C)

Autosampler

  • 6 to 51°C incubation temperature
  • 480 NMR tubes per batch
  • Standard NMR tubes hold 600 µL to 1.5 mL
  • Micro NMR tubes hold 40 µL

2D Experiment Types

  • Correlation spectroscopy (COSY)
  • Double-quantum filtered COSY
  • COSY featuring a 45 degree pulse
  • Total correlation spectroscopy (TOCSY)
  • Heteronuclear single quantum correlation spectroscopy (HSQC, 1H-13C or 1H-15N)
  • Heteronuclear multiple quantum correlation spectroscopy (HMQC)
  • Heteronuclear multiple-bond correlation spectroscopy (HMBC, 1H-13C or 1H-15N)
  • HSQC-TOCSY (1H-13C)
  • HMQC-TOCSY (1H-13C)
  • Nuclear Overhauser effect spectroscopy (NOESY)
  • Rotating-frame Overhauser effect spectroscopy (ROESY)
Bruker Ascent 500

ExperimentAbsorbanceSpectroscopy[]

Example applications include: Substance Quantification, Substance Identification, Analytical Chemistry, Analytical Biochemistry, Organic Compound Analysis

Microfluidic Spectrometer

Light Source

  • Xenon flash lamp

Absorbance Detection

  • 250 to 750 nm wavelength selection

Microfluidics

  • Fixed path length (0.1 or 0.7 mm) allows for quantification
  • Extremely low volumes required (2 μL)
  • 96 samples per batch
Unchained labs lunatic

ExperimentFluorescenceSpectroscopy[]

Example applications include: Molecular Fluorescence, Fluorescence Properties, Substance Identification, Substance Quantification, Organic Compound Analysis, Analytical Chemistry, Analytical Biochemistry

ExperimentLuminescenceSpectroscopy[]

Example applications include: Protein Structure, Molecular Luminescence, Luminescence Properties, Impurity Detection, Substance Identification, Substance Quantification, Organic Compound Analysis, Analytical Chemistry, Analytical Biochemistry

Multimode Plate Reader w/ Monochromators

Excitation

  • Xenon flash lamp
  • 1 to 100 flashes per well
  • 340 to 740 nm excitation monochromator

Emission

  • High voltage PMT detector
  • 340 to 740 nm emission monochromator
  • Top or bottom of plate read mode

Sample Chamber

  • Room temperature (25°C ± 0.45°C (1 sigma)) to 45°C incubation
  • Mix by shaking before read options
  • 1x microtiter plate (6 to 1,536 well) per batch
Clariostar

ExperimentFluorescencePolarization[]

Example applications include: Receptor-Ligand Binding, Enzymatic Degradation, DNA Hybridization

Fluorescence Polarization Plate Reader

  • Xenon flash lamp
  • 1 to 100 flashes per well
  • Dual high voltage PMT detectors
  • 96-well and 384-well plate compatible
  • Excitation Emission Filter Pairs
    • 485 nm, 520 nm
    • 540 nm, 590 nm
    • 625 nm, 675 nm
  • Polarized Dual Simultaneous Emission Filters
Pherastar

ExperimentIRSpectroscopy[]

Example applications include: Receptor-Ligand Binding, Enzymatic Degradation, DNA Hybridization

FTIR Spectrometer

  • Deuterated TriGlycine Sulphate (DTGS) Detector
  • 375 cm⁻¹ to 7,500 cm⁻¹ spectral range
  • > 1,000:1 signal-to-noise ratio
  • > 4 cm⁻¹ spectral resolution
  • Sample amounts from 20 to 70 µL or 20 to 70 mg required
Bruker Alpha

ExperimentDynamicLightScattering[]

Example applications include: Scattered Light Intensity, Brownian Motion, Moving Particles, Particle Size, Polydispersity, Isothermal Stability, Colloidal Stability, Attractive Interaction, Repulsive Interaction

Multimode Spectrophotometer

Multimode Spectrophotometer

  • Measurement in sealed quartz capillaries requires only 9 μL of sample
  • Measure up to 48 samples in one experiment
  • Z-average diameter, polydispersity, isothermal stability, B22, kD, and G22

Dynamic Light Scattering

  • 660 nm laser
  • Avalanche photodiode module scattered light detector
  • 0.3 to 1,000 nm hydrodynamic radius range
  • < 0.1 polydispersity
  • 0.05 mg/mL minimum sample concentration
  • 192 Da to 25 MDa molecular weight range

Thermodynamics

  • Self-contained water-free temperature control system from 15 to 95ºC
  • TEC heating/cooling plate temperature control accuracy 1ºC (at < 70ºC) and 1.5ºC (at > 70ºC)
Unchained lab uncle

DLS Plate Reader

  • 830 nm laser
  • 0.5 to 1000 nm hydrodynamic radius range
  • 1000 to 1 MDa molecular weight range
  • 0.125 mg/mL minimum sample concentration
  • 4 to 85ºC temperature control
  • Support standard 96, 384 or 1,536 microwell plates
  • 4 µL minimum sample volume
Wyatt Dynapro Plate Reader III

ExperimentAbsorbanceKinetics[]

ExperimentFluorescenceKinetics[]

ExperimentFluorescencePolarizationKinetics[]

Example applications include: Protein-Protein Interaction, Live Cellular Monitoring, Reaction Kinetics Analysis, Enzyme Kinetic Measurements, Binding and Dissociation Analysis

ExperimentLuminescenceKinetics[]

Example applications include: Luciferase Reporter Gene Assays, ATP-based Cell Viability Assays, Cytotoxicity Assays, Protein-Protein Interaction, Live Cellular Monitoring, Reaction Kinetics Analysis

All Readers

Sample Chamber

  • Room temperature (25°C ± 0.45°C (1 sigma)) to 45°C incubation
  • Mix by shaking before read options
  • 1x microtiter plate (6 to 1,536 well) per batch

Sample Injection

  • 2x 500 µL injection syringes per batch
  • 50 µL dead volumes, 0.5 µL precision
  • 25 mL/min maximum injection speed
  • 2 mL microcentrifuge tubes, 15 mL conical tubes, or 50 mL conical tube sources
Fluostar Omega

Multimode Plate Reader w/ Monochromators

  • 96-well plate compatible

Excitation

  • Xenon flash lamp
  • 1 to 100 flashes per well
  • 340 to 740 nm excitation monochromator
  • 680 nm laser (AlphaScreen)

Emission

  • High voltage PMT detector
  • 340 to 740 nm emission monochromator
  • Top or bottom of plate read mode

Multimode Plate Reader w/ Filters

  • 96-well plate compatible

Excitation

  • Xenon flash lamp
  • 1 to 100 flashes per well
  • 355 nm, 410 nm, 485 nm, 544 nm, 584 nm, 650 nm, or 700 nm fixed wavelength excitation filters (5 nm bandpass)

Emission

  • High voltage PMT detector
  • 420 nm, 460 nm, 520 nm, 590 nm, 620 nm, or 750 nm fixed wavelength emission filters (5 nm bandpass)
  • Top or bottom of plate read mode

Fluorescence Polarization Plate Reader

Sample Injection

  • 2x 500 μL injection syringes per batch
  • 50 μL dead volumes, 0.5 μL precision
  • 25 mL/min maximum injection speed
  • 2 mL microcentrifuge tubes, 15 mL conical tubes, or 50 mL conical tube sources

Sample Chamber

  • Room temperature (25°C ± 0.45°C (1 sigma)) to 45°C incubation
  • Mix by shaking before read options
  • 1x microtiter plate (6 to 1,536-well) per batch

Optics

  • Xenon flash lamp
  • 1 to 100 flashes per well
  • High voltage PMT detector
  • Top or bottom read mode

Polarization Filters

  • Excitation (vertical), Emission (vertical), Emission (horizontal)
    • 485 nm, 520 nm, 520 nm
    • 540 nm, 590 nm, 590 nm
    • 625 nm, 675 nm, 675 nm

ExperimentNephelometry[]

Nephelometer

Sample Injection

  • 2x 500 µL injection syringes
  • 440 µL/s maximum injection speed

Sample Chamber

  • Room temperature to 65°C incubation
  • Actively regulates O2 and CO2 to 0.1-20%
  • Mix by shaking before read options

Optics

  • 635 ± 10nm 1mW self-monitoring laser diode
  • 1.5 to 3.5 mm selectable beam width
  • Side window photodiode detector
  • Detects up to 80° full cone angle
Nephelostar Plus

ExperimentCircularDichroism[]

Circular Dichroism Plate Reader

  • Dual gratings monochromator
  • 185 to 880 nm wavelength range (2 nm resolution)
  • 96 well plate format
  • 2 minute read time for plate (single wavelength)
  • 1 hour read time for plate (full spectrum)
  • 0.5 L nitrogen gas/min purge
  • 45 µL sample volume
EKKO CD Microplate Reader

ExperimentThermalImaging[]

Thermal Camera

  • Object Temperature Range: -20 to 400°C (-4 to 752°F)
  • Accuracy: ±3°C (at 0 – 100°C) and ±3% (at 100 – 400°C)
  • Resolution: 5 Megapixels
  • Focus: fixed, 0.1m minimum distance
  • Image Frequency: 8.7Hz
  • Spectral Range: 8 to 14µm
Flir

ExperimentNIRImaging[]

ExperimentSWIRImaging[]

ExperimentIRSpectralImaging[]

VIS-SWIR InGaAs Camera

  • High sensitivity digital VIS-SWIR camera
  • Sensor Type: InGaAs PIN-Photodiode
  • Active Pixel: 640 x 512
  • Pixel Pitch: 15µm x 15µm
  • Active Area: 9.6mm x 7.68mm
  • Spectral Response: 0.6 to 1.7µm
  • Readout Noise (rms): < 50 electrons
  • Frame Rate: up to 120Hz
  • Exposure Time: 10µs to 26.8s
  • Onboard Automated Gain Control (AGC)
  • Ultra compact, Low power
Raptor Photonics Owl640

ExperimentNIRFluorsecenceSpectroscopy[]

Fluorometer

  • Signal-to-noise ratio 30,000:1 or better
  • Data acquisition rate: 1,000,000 to 1 point/1000 sec
  • Inputs: 4 photon counting (TTL); 4 analog (+/- 10 volts) ; 1 analog reference channel (+/- 10 volts); 2 TTL
  • Outputs: 2 analog (+/- 10 volts) 2 TTL
  • Emission Range: 185 to 900 nm
  • Light Source: High efficiency “ECO” friendly continuous
  • 75 W Xenon arc lamp
  • Monochromator: Coma-aberration corrected, asymmetrical, excitation or emission optimized, Czerny- Turner design
  • Focal Length: 350 mm
  • Slits: Computer controlled, continuously adjustable
  • Emission Grating: 1200 line/mm 400 nm blaze
  • Wavelength Accuracy: +/- 0.3 nm
  • Minimum Step Size: 0.022 nm
  • Detection: Multimode – Photon Counting, 3 analog (fast, medium, slow response), direct and Single-Shot Transient Digitizer (SSTD) mode
Horiba 8000